SONIMIX 2130
Nox converter tester
The Sonimix 2130 NOx converter tester has been designed to generate a stable NO2 concentration with a small reaction time. It is based on the sonic nozzles technology (ISO 6145-6 technology) and high stability pressure regulators.
The NOx converter tester is often used together with a multi-steps gas divider also based on sonic nozzles and high precision pressure regulators.
In order to validate NOx measurements, the efficiency of the NO2 converter must be higher than 90 %, it is therefore very important to check easily and precisely the efficiency of the NO2 converter.
The SONIMIX 2130 converter tester is based on the process known under the name of Gaseous Phase Titration
Productos Destacados
- Easy use and installation
- US EPA Method
Principales Aplicaciones
- Test of the converter’sefficiency
- Emission measurements on the stack monitoring and automotive industry
Especificaciones Ténicas
Models: SX 2130 | 6800 50 001 | 6800 50 002 | 6800 50 003 |
Gas generation | |||
Concentration | Low | High | |
NO2 & O3 generation | 15 to 50 ppm | 30 to 100 ppm | 300 to 1000 ppm |
Inlet pressure | 2,5 to 4 bar rel.(37 to 58 PSI) | ||
Outlet pressure | Up to 910mbar rel. (13 PSI) | ||
Gas dilution | 90% of the NO cylinder | ||
Stabilization time | NO: < 1 min
O3 and NO2 : < 10 min |
NO: < 1 min
O3 and NO2 : < 10 min |
|
Flow (Air line) | 0,2 Nl/min | ||
Flow (NO line) | 2 Nl/min | ||
Connections | 1/4’’ Swagelok | ||
Materials in contact with gases | NO, O3, NO2: Stainless steel, Teflon, Gold, Silver
Air: Stainless steel, Aluminium, Silver, Brass |
||
Alimentation | 230V / 50-60 HZ
115V / 60HZ |
Solicitar información
Need more information about this product?
Please contact us at info@lni-swissgas.eu or fill the form.