OPT.H2.SENS.2

H2 sensor kit for GC oven

OPT.H2.SENS.2 is a security system that constantly detects the amount of Hydrogen present in the GC’s oven and blocks the Hydrogen generation if it detects an amount greater than the safety threshold.

The OPT.H2.SENS.RVC controller reads the percentage of Hydrogen measured by the sensor OPT.H2.SENS.2 and, if the value exceeds the threshold alarm, the Controller deactivates the solenoid valve in the OPT.H2.SENS.VB.V2 block.
The alarm can be signaled to a PC and to a remote device.

OPT.H2.SENS.2 is engineered to work in standalone mode connected by RS-485 network to a Hydrogen Generator. The sensor can be connected directly with the LNI Hydrogen Generators.

When the hydrogen concentration detected exceeds a threshold of safety, the Hydrogen Generator automatically stops the production and audible and visual alarms will be reported.

Productos Destacados

  • Continuous monitoring of the working environment
  • Compatible with all GCs
  • Simple installation
  • Simple calibration
  • Acoustic and optical alarms
  • Potential explosion risks eliminated
  • Low operating costs

Principales Aplicaciones

  • GC’s oven
  • All environments where Hydrogen is used

Especificaciones Ténicas

Models: OPT.H2.SENS.RV 
OPT.H2.SENS.RVC
Power Supply Voltage 12 VDC +/- 5%
Power Supply Current 0.8 A use with OPT.H2.SENS.VB.V2
Valve output Open collector
12VDC/500mA
Ambient temperature 5-35°C (41-95°F)
Ambient Humidity (max, non condensing) 80% at 25°C (77°F)
IP rating IP20
OPT.H2.SENS.VB.V2
Hydrogen Inlet pressure 1 to 9 bar
Inert Inlet pressure 1 to 10 bar
Pressure regulator 1 to 9 bar

 

Models: OPT.H2.SENS.2
Hydrogen Sensitivity Range 0.01% to 1.00% by hydrogen volume (25% LEL, Lower Explosion Limit)
Accuracy ± (5.0% of F.S. + 1000ppm)
Typical Response Time less than 20 sec
Power Supply Voltage From 9VDC to 28VDC
Communication port RS-485
Protocol MODBUS-RTU
Calibration Interval 1 year (suggested 180 days)
Hydrogen Sensor Technology Dioxide semiconductor
Product Life Expectancy 10 years
Body temperature range -10 °C to 40 °C

 

Solicitar información

Need more information about this product?
Please contact us at info@lni-swissgas.eu or fill the form.

Do you need technical support?

Ask for assistance

Subscribe to our newsletter

Register and receive updates on news, offers and promotions of LNI Swissgas.