NG CASTORE MP 4000

Gas station for Agilent hybrid system pneumatic controlled with tanks

The NitroGen series CASTORE MP 4000 are Membrane Nitrogen Generators. They are engineered to transform standard compressed air into a safe and regulated supply of Nitrogen with a purity up to 99.9%.
An additional channel can provide up to 40 L/min of compressed, dry and regulated air.
The CASTORE MP 4000 with its dual Nitrogen and Air outlets has been developed to meet specific requirements in term of gas, flow, purity and pressure for Agilent MP AES 4xxx series. It can be used for the evaporation of solvents in samples being analysed.

Gas station NG MP4000

Product features

  • Nitrogen + Air generator for Agilent MP AES 4xxx
  • Compatible with all LCMS
  • Dual Nitrogen and Air generator
  • Low noise with no mechanical moving parts
  • Low maintenance
  • Generates nitrogen on demand
  • Designed to run 24 hours a day

Main applications

  • Agilent MP AES 4xxxx
  • LCMS
  • Solvent evaporation

Technical specifications

Model: NG CASTORE MP 4000
N2 outlet
Flow rate (Max) 30 l/min
Outlet pressure (min / Max) 2.0 bars / Inlet pressure – 1.0 bar (14.5 psi)
Nitrogen purity

Inlet Pressure >= 8 bar (116 psi)

> 99.9 % @ 25 l/min

> 99.7 % @ 30 l/min

Internal tank capacity 10 L
Dry Air outlet
Flow rate (Max) 40 l/min
Outlet pressure Inlet pressure – 1.0 bar (14.5 psi)
Dew point *1 < -30°C (-22°F)
Internal tank capacity 10 L
Air inlet
Air inlet requirement ISO 8573-1:2010 Class [2:4:1]
Flow rate (Max) 170 l/min
Supply pressure (min / Max) 7 bars (101 psi) / 10 bars (145 psi)
Temperature (min / Max) 1°C (34°F);  45°C(113°F)
General data
Supply rating 100-240Vac (±10%) 50/60 Hz
Connection type IEC320-C14
Nominal power (Max) 25W (30VA)
Fuse rating (5x20mm) 1A (250Vac – T)
Net weight < 40 kg
Dimensions (W x D x H) 48 x 35 x 120 cm
Connections
N2 outlet port ¼” BSPP female
Dry Air outlet port ¼” BSPP female
Inlet port ¼” BSPP female
Drain port ¼” BSPP female
Operating/storage conditions
Temperature 5-35°C (41-95°F)*2
Humidity (max, non condensing) 80% [5-35°C (41-95°F)]
Noise < 35dB(A)
IP rating IP20
Pollution degree rating 2 (with no aromatic compounds)
Altitude up to 4000m
*1 – Min temperature > 10°C to have best performance on O2 residual content