SONIMIX 3012/3022
Multigas calibrator with GPT, ozone & self re-generable zero air generator
The multigas calibrator with GPT, ozone and self re-generable zero air generator, Sonimix 3012/3022, is an instrument designed to calibrate manually or by remote control up to 6 monitors installed in an air pollution monitoring station, such as SO2, HC, BTX, CO, NOx and O3.
The Sonimix 3012/3022 is equipped with
– an inlet for high concentration mixtures of diluted gases such as NO, NO2, SO2, CO, HC, BTX or mixtures of several compatible polluants in the same cylinder.
– a zero air generator composed of an internal compressor with self re-generable filters that removes moisture, O3, NOx, SO2, HC, BTX and optionally CO and CH4.
– a binary gas dilutor with two inlets (diluted gas and air), based on sonic nozzles (ISO 6145-6 standard)
The ozone generator is regulated by UV-light feedback and compensated from the atmospheric pressure and ambient temperature variations, with alarm on the drift of UV-light as well as the circuits for the GPT.
Produkteigenschaften
- Accuracy – (0,6% rel.)
- Best quality price ratio
- No maintenance nor pressurised Air cylinder with its included self re-generable zero air generator
- Sonic nozzle technology
- Remote control
- Ergonomic with LCD display
Hauptanwendungen
- Air pollution monitoring
Technische Daten
Models: Sonimix 3012/3022 | Sx 3012/3022 | Sx 3012/3022
010 |
General data | ||
Dilution point | 3 | 10 |
Dilution range | 1/166, 1/250, 1/500
(100, 200, 300 ppb with 50 ppm at inlet) |
1/200 to 1/2000 or 1/100 to 1/1000 |
Dilution principle | Sonic nozzle | |
Dilution gas line flow accuracy | 0,6% relative | |
Repeatability on flow | Better than 0.3% relative | |
Warm up time | < 45 min | |
Stabilization time | 3 min | |
Operating temperature | 15°C – 40°C | |
Dimensions | 19″ 3HE/84 TE , 500mm deep
portable casing in option |
|
Net weight | 20 Kg | |
Gas inlet requirments | ||
Zero Air | Requires high grade zero air from internal zero air generator or from cylinder
DP < -40°C |
|
Carrier gas | NO, NO2, O3, SO2 < 1ppb
BTEX < 1 ppb CH4 < 50 ppb & CO HCnm < 50 ppb if catalysis oven is installed |
|
Dilution gas | Requires High concentrated premixtures
SO2 in Air – NO in N2 – CO in Air – HC or BTX in Air |
|
Pressure | 3 ± 0.3 bar rel | |
Fittings | 1/4’’ and 1/8’’ Swagelok | |
Gas outlet requirments | ||
Flow | About 3000 Nml/min, other values upon request | |
Fittings | 1/4’’ and 1/8’’ Swagelok | |
O3 section | ||
O3 concentration | 25 – 500 ppb, adjustable by software | |
Accuracy | Better than +/- 2 ppb from 25 to 200 ppb | |
Repeatability on flow | Better than +/- 2.5 ppb (at 250 ppb O3) | |
Communication | ||
Key board and VFD display | Standard | |
RS232 | Standard | |
Electrical data | ||
Power supply voltage | 230V/50-60Hz
115V/60Hz 100V/50-60Hz |
|
Installed power (Max) | 65 – 150W |
Informationen anfordern
Benötigen Sie weitere Informationen über dieses Produkt?
Bitte kontaktieren Sie uns unter info@lni-swissgas.de oder füllen Sie das Formular aus.