SONIMIX 4001

Ozone generator – Primary transfer standard

The Sonimix 4001 is a stand alone ozone primary transfer standard able to generate ozone together with dry or wet air (optional) allowing to calibrate the ozone monitors in several points. It generates an outflow of 3 Nl/min air containing 10 to 1000 ppb O3, with stability and reproducibility of the concentration better than 1% relative. The extended range model generates 10 ppb to 10 ppm at 3 L/min and 10 ppm to 35 ppm at 0.9 L/min.

The Sonimix 4001 is using pressure regulator and sonic nozzles and an ozone generator including thermostatisation and regulation by UV light photometer measurement with atmospheric pressure and temperature corrections (US EPA method of UV absorption by ozone, Beer-Lambert law). The ozone analyser is built with 1 reference UV light detectors that measure in alternance the UV intensity without any ozone and then the same UV light but through the generated ozone. With these 2 values, the processor calculates the amount of ozone generated.

Ozone generator

Produkteigenschaften

  • Stable and accurate > 1% (rel.)
  • Calibration with dry & wet air
  • Included novel LED photodetector
  • Ergonomic with LCD display

Hauptanwendungen

  • Air pollution monitoring
  • Laboratory (ozone transfer)

Technische Daten

Models:

Sonimix 4001

6800 17 000

standard range

6800 17 001

extended range

General data    
Range From 10 to 1000 ppb in 1 ppb steps @ 3 L/min From 10ppb to 10 ppm @ 3 L/min + 10ppm to 35 ppm @ 0.9 L/min
Dilution principle Sonic nozzles and high precision pressure regulators
Dilution air Dry air from internal zero air (option) or external dry air source
Generated gas flow 3000 ml/min for range 1 900ml/min for range 2
Accuracy 10 ppb to 100 ppb : +/- 2 ppb

>100 ppb : 1.5% relative

Ozone Repeatability Better than 1% relative
Warm up time < 45 min
Stabilization time < 15 min
O3 regulation Measure of the UV light absorption of the generated ozone by UV reference photometer. (US EPA method and Beer Lambert law). Including atmospheric pressure and room temperature corrections
Dimensions 19cmx36cm, 50cm deep
Net weight 15 Kg
UV Section    
UV lamp life time About 10’000 hours (with a mean concentration of 1 ppm)
Zero Air section  
Life Time Annual preventive maintenance
Quality O3, NO and NO2 < 1 ppb       –     H2O < -40°C dew point
Communication    
Outlet Port Teflon 1/4‘’ Swagelok fitting
Electrical data    
Power supply voltage 230V/50-60Hz – 117V/60Hz – 100V/60Hz

 

Informationen anfordern

Benötigen Sie weitere Informationen über dieses Produkt?
Bitte kontaktieren Sie uns unter info@lni-swissgas.de oder füllen Sie das Formular aus.

Benötigen Sie technische Unterstützung?

Service-Anfrage

Erhalten Sie unseren newsletter

Registrieren Sie sich um aktuelle Neuigkeiten, Angebote und Rabattaktionen zu erhalten.